This meter is an insulation resistance tester with broad range of measurement which can be applied to various situations from low-voltage to high-voltage measurements. For example, measuring the insulation of high-voltage receiving station and transformer substation, as well as the electric motor and cable.

  • Ambient temp. and humidity for test: 0 – 40 °C, < 85% RH (without condensation)
    • Temperature for battery charging: 10 – 40 °C, < 80% RH
    • Storage temperature and humidity: -10 – 50 °C, < 90% RH (without condensation)
    • Temperature and humidity for battery pack storage: -20 – 30 °C, < 80% RH (without condensation) • Operation environment: For elevation lower than 2000 m • Test method: Applying DC voltage (for measuring insulation resistance); effective value (voltage) • Display: LCD with backlight; Max. number: 999 • Overflow indication: >, OF
    • Underflow indication: >, -OF
    • Frequency for updating the display: For insulation resistance / leak current: once per second (once in every four seconds if AVERAGE function is used )
    • Detection of output voltage: Twice per second
    • Voltage: Four times per second
    • Temperature: Once per second
    • Graduation bar: Twice per second
    • Terminals:
    1. Insulation resistance / voltage test: L(+), E(-), GUARD (GUARD terminals
    can only be used for insulation resistance / leak current tests).
    2. Temperature sensor, USB, and charger.
    • Power supply:
    1. LR14 alkaline battery × 6; rated voltage: 1.5 V × 6
    2. Battery pack, rechargeable nickel – metal hydride batteries; rated voltage: 7.2 V (for 500 times of charging)
    • Max. power consumption:
    1. 10 VA (using battery);
    2. 6 VA (using rechargeable battery pack).
    • Max. powering time: Alkaline battery: about 9 hours; Battery pack: about 5 hours (with 5 kV-voltage generated, with terminals in an open circuit, and with backlight turned off).
    • Max. input voltage: 750 VAC, ± 1000 VDC
    • Max. input frequency: 70Hz
    • Max. rated voltage to earth: 1000 Vrms (CAT III), 600 Vrms (CAT IV)
    • Insulation strength: 6880 VAC: 15 seconds
    • Overload protection: 1000 VAC, 1200 VDC between L(+) and E(-) terminals: 1 minute
    • Dimensions: About 284 mm (L)×230 mm (w)×125 mm (h)
    • Net Weight: About 2.5 kg (without batteries)
    • Applicable standards: The design is complied with DL/T 845.1-2004
    EN61010-1:2001
    • Interface: USB Ver 2.0, for communications with the PC software.
    • PC application software:
    1. Data transfer
    2.Tester setup
    3. Data output